What are the best practices for loading samples onto the SeekOne DD chip?

2026-06-23

Answer:
Strict adherence to loading order and technique is critical to ensure stable droplet generation and reproducible results.

Required loading order:

  1. Row 1: Cell suspension
    (Prepare the mix first; add cells last) 
  2. Row 2: Barcoded beads
    (Equilibrate to room temperature for 30 min; vortex thoroughly before use) 
  3. Row 3: Carrier oil 

Key operational guidelines:

  • When dispensing reagents, keep the pipette tip ~3 mm below the liquid surface to avoid introducing air bubbles. 
  • Barcoded beads are viscous:
    After aspirating to the target volume, wait ~5 seconds before removing the tip to ensure accurate volume transfer. 
  • Allow a ~30-second interval between loading each well
  • For multiple chips:
    Load by row across all chips (all cells → all beads → all oil), rather than completing one chip at a time. 
  • During preparation of the cell mixture:
    Add cells last and mix gently but thoroughly before loading. 

After loading all wells:
Ensure no visible bubbles remain at the bottom of the wells.

 