What should be done if residual reagents or oil are observed on the SeekOne DD Chip Holder after droplet generation?

2026-06-18

Answer:
After completion of droplet generation, residual reagents may remain on the SeekOne DD Chip Holder.

During operation of the SeekOne DD Chip S3, a relatively large volume of carrier oil is loaded into the chip. If the chip holder is subjected to movement, impact, or is not maintained in a level position, there is a risk that oil or reagents may spill onto the SeekOne DD Chip Holder.

In such cases, the impact on experimental performance should be evaluated based on the recovered emulsion:

  • If the collected emulsion from Chip S3 is uniform in appearance and the total volume is approximately 100–120 μL, the assay performance is not expected to be affected. 
  • If the emulsion remains uniform but the volume is slightly reduced, data quality is generally not impacted. However, if there is any uncertainty, it is recommended to contact technical support for further evaluation.  

If oil residue is present on the chip holder, it should be cleaned as follows:

  • Wipe the surface using Kimwipes in combination with isopropanol (IPA), either as pre-moistened wipes or applied via a spray bottle. 
  • Ensure that all visible oil residues are removed to maintain cleanliness and prevent contamination. 

To prevent recurrence, the following precautions are recommended:

  • Always keep the assembled chip holder in a horizontal (level) position. 
  • Handle the chip holder carefully during movement, placement, or opening. 

Avoid wetting the gasket or allowing reagents to spill into the wells, as this may affect subsequent runs.

 